Transform 680 Microwave Digestion System

The Transform 680 is a top-loading, closed vessel microwave digestion system with high-pressure capabilities to enhance digestion quality while decreasing digestion time. This system is capable of simultaneously running up to 6 high-pressure closed vessels, providing a fast and automated method to digest even the most difficult samples. With simple one-step operation and full computer control of temperature and power, this cost-effective microwave digestion system consolidates your AA, ICP-AES, and CP-MS applications. The MW 680 is a cost-effective system that provides optimal performance while leading the industry in safety

 Download Transform 680 Brochure
 104 KB







Features

Your MW 680 Closed Vessel System includes:
• Round, pressure resistant heavy duty oven chamber made of PFA-coated stainless steel
• Evenly distributed microwave field throughout the chamber to ensure uniform heating for each sample vessel.
• 6 closed vessels designed to operate at up to 250°C temperature and 400psi pressure.
• Maximum 1200 watts microwave output with fast, seamless power control for the full power range.
• A sealed pressure control system.
• A microwave-immune thermocouple temperature control probe.
• Vessels equipped with Teflon® inert inner surface and Ultem® outer covering for superior mechanical strength.
• Ever present, operator transparent, electronic temperature safety interlocks.
• Over-pressure venting safety interlock.
• Top-loading door push-button with mechanical safety interlocks on the oven door
• Separate dual exhaust systems - one for the electronics and the other for the microwave oven cavity.
• Adaptive control software, that allows total controlled programming of temperature independent of the workload. (Support      Windows 2000 and Windows XP operating system)
• Lifetime applications support from MW Discovery.

Closed Vessel Digestion
Closed vessel digestion reduces contamination issues by ensuring a clean, controlled environment for each individual sample. The MW 680 can hold 6 high-pressure vessels operating at 400 psi and 250°C. A sealed thermocouple sensor enable fast and accurate temperature control.

Safety
The Transform 680 comes standard with a range of safety features:
• Top-loading
• Fume exhaust module
• Rupture disks and vent detection mechanisms on each vessel automatically shut down power in response to 
    over-pressure
• Teflon lining of oven cavity
• Digestion vessels molded with engineering plastics
• Continuous temperature monitoring and control
• Release of acid fumes into collection reservoir in vessel carousel
• Automatic pressure release mechanism in oven cavity

Computer Control
Transform 680 is a full computer controlled system. Its software features direct control and sensing of temperature with real-time graphic display, automatic data storage, unlimited method storage, and complete record of instrument parameters

Specifications

Vessel Module
Number of Vessels: 6 (5 standard+1 sensor vessel)
Operating Pressure: 400 psi
Operating Temperature: 250°C
Temperature Sensor: 1
Vessel Volume: 60 mL

Electrical: 110,60 Hz, 12A & 220V, 60Hz, 20A
Microwave Power: 1200W
Magnetron Frequency: 2450 Mhz
Dimensions: 48 (W) X 46 (D) X 43 (H) cm